Application guide

Instruments

Gas used

Purity Flowrate used Generator model

Gas Chromatographie

GC-FID

Hydrogen

Hydrogen high purity 30 to 50 cc/mn Hydrogen
Zero Air for FID Without hydrocarbon 300 to 500 cc/mn Zero Air
Hydrogen for carrier gas Hydrogen high purity 10 cc/mn Hydrogen
Nitrogen for carrier gas Nitrogen high purity ; N2 Zero 20 to 50 cc/mn Nitrogen NG model  
Nitrogen for make up gas Nitrogen high purity ; N2 Zero 30 to 50 cc/mn Nitrogen NG model  
GC-FPD Hydrogen Hydrogen high purity 60 to 90 cc/mn Hydrogen
Zero Air for FID Without hydrocarbon 90 to 120 cc/mn Zero Air
GC-NPD Hydrogen for carrier gas Hydrogen high purity 50 cc/mn Hydrogen
Nitrogen for make up gas Nitrogen high purity ; NZero 30 to 50 cc/mn Nitrogen NG model  
GC-ECD Nitrogen for carrier gas Nitrogen high purity ; NZero 30 cc/mn Nitrogen + Zero Air
Nitrogen for make up gas Nitrogen high purity ; NZero 60 cc/mn Nitrogen + Zero Air
GC-TCD Hydrogen for carrier gas Hydrogen high purity 100 cc/mn Hydrogen
GC-ATD Dry Air Dry Air < 2 L/mn Dry Air
GC-AED Nitrogen for carrier gas Nitrogen high purity ; NZero < 1 L/mn Nitrogen + Zero Air
GC-ELCD Hydrogen into reacting gas Hydrogen high purity 70 to 200 cc/mn Hydrogen

LC / MS

APCI, ESI Applied Biosystem Air and Nitrogen for nebulisation Nitrogen / Dry Air Until 20 L/mn for Air and 5 L/mn for Nitrogen Nitrogen NMG  and  Dry air
LCMS / MS, TOF Nitrogen for curtain gas 99,5 % 5 to 35 L/mn Nitrogen ZEFIRO
Spectroscopy
FT-IR Purge gas for sample compartment, optics, air bearing and microscope Dry Air without CO2 15 to 85 L/mn Dry Air without CO2
RMN Air for lifting spinning and ejecting Dry Air 300 L/mn Dry Air
ICP Nitrogen or Zero Nitrogen for purge gas > 99,995 % 1 to 9 L/mn Nitrogen NG Model
 
AA Air for oxidant gas Dry Air 28 to 200 L/mn Dry Air
Products for Analyzers
COT Dry Air or Nitrogen for carrier gas or combustion gas Dry Air without CO2, without hydrocarbon 100 to 500 cc/mn Zero Air + Dryer without CO2
Nitrogen high purity 50 to 700 cc/mn Nitrogen NG model
Analyseur CO2 Calibration Air Dry Air without CO2, without hydrocarbon 550 to 1000 cc/mn Ultra Zero Air
DSC Air for air shield Dry Air 100 cc/mn Dry Air
TGA Nitrogen or Dry Air as furnace gas Nitrogen high purity 100 cc/mn Dry Air or Nitrogen NG model
Other lab applications
Sample prep Nitrogen for solvent evaporation 95 % à 99 % 130 L/mn Nitrogen NG model
Air for pneumatic control Dry Air 28 L/mn Dry Air
Nitrogen for auto-sampler Nitrogen high purity, Nitrogen Zero 550 cc/mn Nitrogen NG model
Circular Dichroïsm Nitrogen 98 % 2 to 8 L/mn Nitrogen N2
ELSD Detector Air for nebulisation Dry Air or Nitrogen 2 to 6 L/mn Dry Air, Nitrogen ELSD model
Particle sizing by laser diffraction Clean and dry air for nebulizing  gas Dry Air - Dry Air

 

Air quality of compressed air  FOR NITROGEN generators for LCMS, and zero air :

 

ISO 8573.1 CLASS 1.2.1

 

 

 

The new international standard for compressed air quality introduces a simple classification system for the 3 main contaminants present in any compressed air system : dirt, water and oil.

 

Quality class

Dirt Particules size in micron

Water Pressure dewpoint °C at 7 bar

Oil Including vapor mg/m3

1 0,1 -70°C 0,01
2 5 -40°C 0,1
3 10 -20°C 1
4 40 3°C 5
5 - 7°C 25
6 - 10°C -

 

Note :

 

If the LCMS nitrogen generator is connected to a non-lubrificated compressor, the fridge dryer is efficient.

 

If the compressor is lubrified, it is recommended to use an adsorption dryer.